Description
The HI-2020 Edge® measures pH, conductivity, and dissolved oxygen through its digital electrodes. These digital electrodes are auto-recognised, providing sensor type, calibration data and a serial number when connected to edge® (with plug-in 3.5mm connector).
The electrode holder features a swivel, adjustable arm with built-in cradle to hold edge pH meter securely in place at optimum viewing angle.
Supplied with bench-top stand; electrode holder; bracket for wall-mount; USB cable; power adaptor; instruction manual.
edge® pH kit: HI-11310 digital pH electrode with integrated temperature sensor; sachets of pH 4.01, 7.01, 10.01 buffer solutions; sachet of electrode cleaning solution.
Optional accessories include:
edge® DO kit: HI-764080 dissolved oxygen electrode, HI7041S refill electrolyte solution, DO membrane caps, o-rings.
edge® EC kit: HI-763100 conductivity probe; 1413, 12880, 5000 µS/cm conductivity standard sachets; electrode rinse solution sachets.
Specification
pH Measurement:
Basic & Standard Mode: – 2.00 to 16.00 pH; -2.000 to 16.000 pH (standard mode)
Resolution: 0.01 pH, 0.001 pH (standard mode)
Accuracy (at 25°C): ± 0.01 pH, ± 0.002 pH (standard mode)
pH Calibration:
5 points (standard mode) and two custom buffers: 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45
3 points (basic mode): 4.01, 6.86, 7.01, 9.18, 10.01
pH Temperature Compensation: ATC: -5 to 100°C; 23 to 212°F
mV pH Measurement:
Range: ±1000 mV
Resolution: 0.1 mV
Accuracy (at 25°C): ±0.2 mV
EC Measurement:
Range: 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm absolute EC (with temperature compensation function disabled)
Resolution: 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm
Accuracy (at 25°C): ±1% of reading (±0.5 µS or 1 digit, whichever is greater)
TDS Measurement:
Range: 0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L absolute TDS using 0.80 conversion factor (with temperature compensation function disabled)
Resolution: 0.01 mg/L (ppm); 0.1 mg/L (ppm); 1 (ppm); 0.01 g/L; 0.1 g/L
Accuracy (at 25°C): ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
DO Measurement:
Range: 0.00 to 45.00 ppm (mg/L); 0.0 to 300.0 % saturation
Resolution: 0.01 ppm (mg/L); 0.1 % saturation
Accuracy (at 25°C): ± 1.5% of reading ±1 digit
Temperature Measurement:
Range: -20 to 120°C, -4 to 248°F
Resolution: 0.1°C, 0.1°F
Accuracy: ± 0.5°C, ± 0.9°F
GLP: Yes
Logging: Stores up to 1000 records: 200 records (log-on-demand and stability logging)
Connectivity: 1 micro USB port for charging and PC connectivity; 1 USB port for storage
Environment: 0 to 50°C, RH max. 95% non-condensing
Power Supply: 5 VDC adapter (included)
Dimensions: 202 x 140 x 12mm
Weight: 250g