Description
The HI-2004 Edge® dissolved oxygen meter covers all ranges from 0.00 to 45.00ppm (mg/L); 0.0 to 300.0% saturation with an accuracy of ±1.5%.
Supplied with bench-top stand; electrode holder; bracket for wall-mount; USB cable; power adaptor; instruction manual. HI-764080 dissolved oxygen electrode, HI-7041S refill electrolyte solution, DO membrane caps, o-rings.
Specification
DO Measurement:
Range: 0.00 to 45.00ppm (mg/L); 0.0 to 300.0% saturation
Resolution: 0.01ppm (mg/L); 0.1% saturation
Accuracy (at 25°C): ± 1.5% of reading ±1 digit
Calibration: One or two-point at 0% (HI7040 solution) and 100% (in air)
Temperature Compensation: ATC (0 to 50°C; 32.0 to 122.0°F)*
Salinity Compensation: 0 to 40g/L (with 1 g/L resolution)
Altitude Compensation: -500 to 4000m (-1640 to 13120′) (with 100m (328′) resolution)
* temperature limits will be reduced to actual probe limits
Temperature:
Range: -20 to 120°C, -4 to 248°F
Resolution: 0.1°C, 0.1°F
Accuracy: ± 0.5°C, ± 0.9°F
GLP: Yes
Logging: Stores up to 1000 records: 200 records (log-on-demand and stability logging)
Connectivity: 1 micro USB port for charging and PC connectivity; 1 USB port for storage
Environment: 0 to 50°C, RH max. 95% non-condensing
Power Supply: 5 VDC adapter (included)
Dimensions: 202 x 140 x 12mm
Weight: 250g