Description
Priorclave’s Smart range adeptly handles standard tasks (e.g. media prep, waste loads, labware) but can also produce drier loads and more reliably handle complex loads that tend to trap air. Ideal for mixed waste, biological waste, pathogen-heavy materials, cloth and textile sterilisation, soil and substrate preparation, wrapped instruments, and porous loads.
These sterilisers feature fully customisable cycles (with expanded memory), pre-cycle and post-cycle vacuum, integrated cooling and free-steaming, and air intake filtration.
These smaller top-loading sterilisers a popular choice, owing to their high efficiency and extremely small footprint (relative to chamber size).
Features
• Quickseal single-action door fitted with thermal and pressure locks to prevent opening at load temperature above 80°C and pressures above 0.2 bar
• Tactrol® 3 microprocessor control with simple and fully variable setting of process time, temperature with graphic indication of cycle status
• Automatic timed free-steaming
• Forced air cooling
• Pre-cycle vacuum in which multiple Vacuum stages are programmed with interspersed heating to achieve steam penetration
• Pulsed free-steaming to maximise efficiency of Pre-cycle vacuum phase
• Media Warming and Delayed Start functions keep autoclaved media at a usable temperature ready for immediate use
• Vacuum assisted cooling
• Low water level sensor
• Two or three loading baskets
• Epoxy coated panels and frame members are treated with an anti-bacterial agent which is effective against all bacteria and fungi including MRSA
• Pressure vessel carries a 10 year warranty and is insurance approved
• Service and support direct from manufacturer (or appointed distributor)
Specification
Autoclave Capacity (litres):
60
Chamber Size:
350 x 625mm
Chamber Material:
Polished 316 stainless steel
Loading Format:
Front
Operating Range:
Up to 138°C, 2.4 bar
Heater Power:
3kW at 13A single phase
Water Supply:
N/A
Weight (kg):
90
External Dimensions (w x d x h):
543 x 910 x 915mm
Installation Dimensions:
Please allow 500mm (19.69”) either side and 300mm (11.81”) at the back of the autoclave